Design and application of multiparameters contrast detail phantom using cylindrical double hole acrylic block fror radiographic quality control tool /

The consistency of contrast detail performance of imaging system can be evaluated by using Contrast Detail phantom. The analysis is always based on the human visual perception which leads to intraobserver perception bias. In addition, the wall of single drilled hole concept in commercial Contrast De...

Full description

Saved in:
Bibliographic Details
Main Author: Nur Farahana binti Pauzi
Format: Thesis
Language:English
Published: Kuantan, Pahang : Kulliyyah of Allied Health Sciences, International Islamic University Malaysia, 2016
Subjects:
Online Access:Click here to view 1st 24 pages of the thesis. Members can view fulltext at the specified PCs in the library.
Tags: Add Tag
No Tags, Be the first to tag this record!
LEADER 031260000a22002770004500
008 171114t2016 my a g m 000 0 eng d
040 |a UIAM  |b eng 
041 |a eng 
043 |a a-my--- 
050 |a TK8315 
100 0 |a Nur Farahana binti Pauzi 
245 1 |a Design and application of multiparameters contrast detail phantom using cylindrical double hole acrylic block fror radiographic quality control tool /  |c by Nur Farahana binti Pauzi 
260 |a Kuantan, Pahang :  |b Kulliyyah of Allied Health Sciences, International Islamic University Malaysia,  |c 2016 
300 |a xxiv, 159 leaves :  |b ill. ;  |c 30cm. 
502 |a Thesis (MHSC)--International Islamic University Malaysia, 2016. 
504 |a Includes bibliographical references (leaves 150-155). 
520 |a The consistency of contrast detail performance of imaging system can be evaluated by using Contrast Detail phantom. The analysis is always based on the human visual perception which leads to intraobserver perception bias. In addition, the wall of single drilled hole concept in commercial Contrast Detail phantom gives effect to the penetration of x -ray beam divergence to pass through the base of each hole. This effect will lead to false appearance of image but it is not visualised in the radiograph. In this study, cylindrical double hole acrylic block of Multiparameters Contrast Detail phantom has been developed which differs from the single drilled hole concept, whereby it consists of combination of different holes' diameters and thicknesses. Results revealed that, the new design of cylindrical double hole acrylic block is able to visualise the effect of image displacement from the x-ray focal length plus measuring the off-position of anode stem, blurring effect, image distortion in terms of real shape and size, also addressing the contrast detail characteristics parameter in terms of real hole depth. The influencing factors of source-image-distance, object thickness, position of object from center beam in x-axis, sizes of hole diameter and hole depth contribute to the changes in parameters' outcome. The measurement of pixel intensity by using software and development of algorithm for data analysis basically can reduce the human perception bias and increase the validity of the results. 
596 |a 1 6 
655 7 |a Theses, IIUM local 
690 |a Dissertations, Academic  |x Department of Radiography & Diagnostic Imaging  |z IIUM 
710 2 |a International Islamic University Malaysia.  |b Department of Radiography & Diagnostic Imaging 
856 4 |u https://lib.iium.edu.my/mom/services/mom/document/getFile/ZbPtN24gjsOaL1ObjO31at76rN6CD70K20161124112747054  |z Click here to view 1st 24 pages of the thesis. Members can view fulltext at the specified PCs in the library. 
900 |a sbh-lfr 
999 |c 436223  |d 467789 
952 |0 0  |6 T TK 008315 N974D 2016  |7 0  |8 THESES  |9 761208  |a KIMC  |b KIMC  |c CLOSEACCES  |g 0.00  |o t TK 8315 N974D 2016  |p 11100347313  |r 2019-09-04  |t 1  |v 0.00  |y THESIS 
952 |0 0  |6 TS CDF TK 8315 N974D 2016  |7 0  |8 THESES  |9 854802  |a IIUM  |b IIUM  |c MULTIMEDIA  |g 0.00  |o ts cdf TK 8315 N974D 2016  |p 11100347314  |r 2018-02-06  |t 1  |v 0.00  |y THESISDIG