Effects of prosthesis stem length on stress distribution in cementless hip arthroplasty / Emmi Farisa Jaafar.

Total Hip Replacement (THR) represents a modern surgical technique that allow for replacement of the natural hip joint by an artificial one. In this study, finite element analysis (FEA) of the THR is performed. The effects of prosthesis stem lengths on the resulting stress are obtained. Three differ...

Full description

Saved in:
Bibliographic Details
Main Author: Jaafar, Emmi Farisa
Format: Thesis
Language:English
Published: 2009
Subjects:
Online Access:https://ir.uitm.edu.my/id/eprint/31149/1/31149.pdf
Tags: Add Tag
No Tags, Be the first to tag this record!
id my-uitm-ir.31149
record_format uketd_dc
spelling my-uitm-ir.311492020-06-15T03:22:54Z Effects of prosthesis stem length on stress distribution in cementless hip arthroplasty / Emmi Farisa Jaafar. 2009 Jaafar, Emmi Farisa Machine tools and machining Machine design and drawing Total Hip Replacement (THR) represents a modern surgical technique that allow for replacement of the natural hip joint by an artificial one. In this study, finite element analysis (FEA) of the THR is performed. The effects of prosthesis stem lengths on the resulting stress are obtained. Three different lengths of prosthesis stem namely short, medium and long are analyzed. The result of each stem are discussed and compared. The prosthesis stem design is based on established model of Charnley prosthesis while the femur model is gathered from net. Both design were modeled and modified using CATIA software while MSC PATRAN, NASTRAN software is used for analysis. The stem is defined as bio compatibility material that is Ti-6AI-4V while the femur is based on cortical bone properties. The loads apply to the THR model are based on physiological load of level walking conditions of human with average weight of 80Kg. Two major loads are considered in the model that are joint contact force and abductor muscle force and fixed at the bottom end. The results are discussed with respect to Von Misses stress of the analysis. Cementless THR with medium stem represent th-e best stress distribution as compare to intact femur. It is suggested to have experimental result for appropriate findings. 2009 Thesis https://ir.uitm.edu.my/id/eprint/31149/ https://ir.uitm.edu.my/id/eprint/31149/1/31149.pdf text en public degree Universiti Teknologi MARA Shah Alam Faculty of Mechanical Engineering Abdullah, Abdul Halim
institution Universiti Teknologi MARA
collection UiTM Institutional Repository
language English
advisor Abdullah, Abdul Halim
topic Machine tools and machining
Machine design and drawing
spellingShingle Machine tools and machining
Machine design and drawing
Jaafar, Emmi Farisa
Effects of prosthesis stem length on stress distribution in cementless hip arthroplasty / Emmi Farisa Jaafar.
description Total Hip Replacement (THR) represents a modern surgical technique that allow for replacement of the natural hip joint by an artificial one. In this study, finite element analysis (FEA) of the THR is performed. The effects of prosthesis stem lengths on the resulting stress are obtained. Three different lengths of prosthesis stem namely short, medium and long are analyzed. The result of each stem are discussed and compared. The prosthesis stem design is based on established model of Charnley prosthesis while the femur model is gathered from net. Both design were modeled and modified using CATIA software while MSC PATRAN, NASTRAN software is used for analysis. The stem is defined as bio compatibility material that is Ti-6AI-4V while the femur is based on cortical bone properties. The loads apply to the THR model are based on physiological load of level walking conditions of human with average weight of 80Kg. Two major loads are considered in the model that are joint contact force and abductor muscle force and fixed at the bottom end. The results are discussed with respect to Von Misses stress of the analysis. Cementless THR with medium stem represent th-e best stress distribution as compare to intact femur. It is suggested to have experimental result for appropriate findings.
format Thesis
qualification_level Bachelor degree
author Jaafar, Emmi Farisa
author_facet Jaafar, Emmi Farisa
author_sort Jaafar, Emmi Farisa
title Effects of prosthesis stem length on stress distribution in cementless hip arthroplasty / Emmi Farisa Jaafar.
title_short Effects of prosthesis stem length on stress distribution in cementless hip arthroplasty / Emmi Farisa Jaafar.
title_full Effects of prosthesis stem length on stress distribution in cementless hip arthroplasty / Emmi Farisa Jaafar.
title_fullStr Effects of prosthesis stem length on stress distribution in cementless hip arthroplasty / Emmi Farisa Jaafar.
title_full_unstemmed Effects of prosthesis stem length on stress distribution in cementless hip arthroplasty / Emmi Farisa Jaafar.
title_sort effects of prosthesis stem length on stress distribution in cementless hip arthroplasty / emmi farisa jaafar.
granting_institution Universiti Teknologi MARA Shah Alam
granting_department Faculty of Mechanical Engineering
publishDate 2009
url https://ir.uitm.edu.my/id/eprint/31149/1/31149.pdf
_version_ 1783734104053776384