Biomechanical analysis of parietal-temporal implant: effects of implant materials, pore shapes and pore sizes / Wan Nur Fatini Syahirah W. Dagang

Cranial implants are designed to shield the brain before and after neurosurgery or trauma, thus their durability is crucial clinical value. The mechanical response of each implant cannot be practically described experimentally due to the rise in patient-specific implants. However, computational mode...

Full description

Saved in:
Bibliographic Details
Main Author: W. Dagang, Wan Nur Fatini Syahirah
Format: Thesis
Language:English
Published: 2023
Subjects:
Online Access:https://ir.uitm.edu.my/id/eprint/88948/1/88948.pdf
Tags: Add Tag
No Tags, Be the first to tag this record!
id my-uitm-ir.88948
record_format uketd_dc
spelling my-uitm-ir.889482024-01-03T06:34:41Z Biomechanical analysis of parietal-temporal implant: effects of implant materials, pore shapes and pore sizes / Wan Nur Fatini Syahirah W. Dagang 2023 W. Dagang, Wan Nur Fatini Syahirah Occupational Health Cranial implants are designed to shield the brain before and after neurosurgery or trauma, thus their durability is crucial clinical value. The mechanical response of each implant cannot be practically described experimentally due to the rise in patient-specific implants. However, computational models such as finite element analysis (FEA) have a great chance of accurately forecasting the mechanical behaviour of implants designed specifically for a certain patient. FEA has been used extensively for the solid cranial implant, yet there has been relatively little research on FEA for porous implants. Thus, this study aimed to analyse the biomechanical behaviour of the solid implant and porous implant by varying implant pore shape (pentagonal and hexagonal), pore size (900 and 1200 μm), and materials (titanium, hydroxyapatite, and zinc-hydroxyapatite) through a finite element method. In the first phase of this research, the three–dimensional (3D) human skull model was developed from the 245 DICOM images with a thickness of 1 mm of a 15-year-old girl patient with a right parietal-temporal defect. The defective skull images underwent the segmentation technique using a threshold method with a Hounsfield unit of 226 HU to 3071 HU via the Mimics software. The excessive bone such as the neck area was removed using a split mask tool and the region of interest was exported as Standard Tessellation Language (STL) file. Next, the 3D skull model was loaded into 3 Matic software. 2023 Thesis https://ir.uitm.edu.my/id/eprint/88948/ https://ir.uitm.edu.my/id/eprint/88948/1/88948.pdf text en public masters Universiti Teknologi MARA (UiTM) College of Engineering Abdullah, Abdul Halim
institution Universiti Teknologi MARA
collection UiTM Institutional Repository
language English
advisor Abdullah, Abdul Halim
topic Occupational Health
spellingShingle Occupational Health
W. Dagang, Wan Nur Fatini Syahirah
Biomechanical analysis of parietal-temporal implant: effects of implant materials, pore shapes and pore sizes / Wan Nur Fatini Syahirah W. Dagang
description Cranial implants are designed to shield the brain before and after neurosurgery or trauma, thus their durability is crucial clinical value. The mechanical response of each implant cannot be practically described experimentally due to the rise in patient-specific implants. However, computational models such as finite element analysis (FEA) have a great chance of accurately forecasting the mechanical behaviour of implants designed specifically for a certain patient. FEA has been used extensively for the solid cranial implant, yet there has been relatively little research on FEA for porous implants. Thus, this study aimed to analyse the biomechanical behaviour of the solid implant and porous implant by varying implant pore shape (pentagonal and hexagonal), pore size (900 and 1200 μm), and materials (titanium, hydroxyapatite, and zinc-hydroxyapatite) through a finite element method. In the first phase of this research, the three–dimensional (3D) human skull model was developed from the 245 DICOM images with a thickness of 1 mm of a 15-year-old girl patient with a right parietal-temporal defect. The defective skull images underwent the segmentation technique using a threshold method with a Hounsfield unit of 226 HU to 3071 HU via the Mimics software. The excessive bone such as the neck area was removed using a split mask tool and the region of interest was exported as Standard Tessellation Language (STL) file. Next, the 3D skull model was loaded into 3 Matic software.
format Thesis
qualification_level Master's degree
author W. Dagang, Wan Nur Fatini Syahirah
author_facet W. Dagang, Wan Nur Fatini Syahirah
author_sort W. Dagang, Wan Nur Fatini Syahirah
title Biomechanical analysis of parietal-temporal implant: effects of implant materials, pore shapes and pore sizes / Wan Nur Fatini Syahirah W. Dagang
title_short Biomechanical analysis of parietal-temporal implant: effects of implant materials, pore shapes and pore sizes / Wan Nur Fatini Syahirah W. Dagang
title_full Biomechanical analysis of parietal-temporal implant: effects of implant materials, pore shapes and pore sizes / Wan Nur Fatini Syahirah W. Dagang
title_fullStr Biomechanical analysis of parietal-temporal implant: effects of implant materials, pore shapes and pore sizes / Wan Nur Fatini Syahirah W. Dagang
title_full_unstemmed Biomechanical analysis of parietal-temporal implant: effects of implant materials, pore shapes and pore sizes / Wan Nur Fatini Syahirah W. Dagang
title_sort biomechanical analysis of parietal-temporal implant: effects of implant materials, pore shapes and pore sizes / wan nur fatini syahirah w. dagang
granting_institution Universiti Teknologi MARA (UiTM)
granting_department College of Engineering
publishDate 2023
url https://ir.uitm.edu.my/id/eprint/88948/1/88948.pdf
_version_ 1794192179794542592