Uniformity, sensitivity, linearity, and hot region image analysis by employing unconventional filter technique (Cu 0.125mm and AI 0.2mm) with Tc-99m inSPECT imaging
In SPECT imaging, the presence of the scattered events in projection data affects the overall quality of the image as well as the quantification of the radioactivity distribution. In this study, material filter method has been employed in the attempt of reducing the influence of scatter events...
Saved in:
Main Author: | |
---|---|
Format: | Thesis |
Language: | English |
Published: |
2004
|
Subjects: | |
Online Access: | http://eprints.usm.my/44154/1/LI...Fong%20Shu%20Tee...2004...-24%20pages.pdf |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
id |
my-usm-ep.44154 |
---|---|
record_format |
uketd_dc |
spelling |
my-usm-ep.441542020-10-22T03:03:26Z Uniformity, sensitivity, linearity, and hot region image analysis by employing unconventional filter technique (Cu 0.125mm and AI 0.2mm) with Tc-99m inSPECT imaging 2004 Vee, Fong Shu R Medicine In SPECT imaging, the presence of the scattered events in projection data affects the overall quality of the image as well as the quantification of the radioactivity distribution. In this study, material filter method has been employed in the attempt of reducing the influence of scatter events in the measured counts. Such a filter decreases the relative contribution of scattered gammas by preferentially removing them before they can reach the surface of the detector of the gamma camera. The material filters used are copper (Cu) sheet 0.125mm thick and aluminium (AI) sheet 0.2mm thick. Filters are separately mounted on the face of a gamma camera (GCA-901 A/HG) with either low energy general purpose (LEGP) or low energy high resolution (LEHR) collimators. SPECT images of the insert simulating 'hot' lesions in cold background, linearity and the uniform section of the Carlson phantom loaded with Tc-99m are obtained without and with material filter. The performance of the tomographic system is explored via the uniformity, sensitivity and linearity measurements, as well as the image lesi9n (hot) contrast, signal-to-noise ratio, standard deviation and percent root mean square noise (PRMSN). From the results, material filtered data shows an improvement in all tested parameters at the cost of system sensitivity. 2004 Thesis http://eprints.usm.my/44154/ http://eprints.usm.my/44154/1/LI...Fong%20Shu%20Tee...2004...-24%20pages.pdf application/pdf en public masters Universiti Sains Malaysia Pusat Pengajian Sains Kesihatan |
institution |
Universiti Sains Malaysia |
collection |
USM Institutional Repository |
language |
English |
topic |
R Medicine |
spellingShingle |
R Medicine Vee, Fong Shu Uniformity, sensitivity, linearity, and hot region image analysis by employing unconventional filter technique (Cu 0.125mm and AI 0.2mm) with Tc-99m inSPECT imaging |
description |
In SPECT imaging, the presence of the scattered events in projection data
affects the overall quality of the image as well as the quantification of the
radioactivity distribution. In this study, material filter method has been employed
in the attempt of reducing the influence of scatter events in the measured counts.
Such a filter decreases the relative contribution of scattered gammas by
preferentially removing them before they can reach the surface of the detector of
the gamma camera. The material filters used are copper (Cu) sheet 0.125mm
thick and aluminium (AI) sheet 0.2mm thick. Filters are separately mounted on
the face of a gamma camera (GCA-901 A/HG) with either low energy general
purpose (LEGP) or low energy high resolution (LEHR) collimators.
SPECT images of the insert simulating 'hot' lesions in cold background,
linearity and the uniform section of the Carlson phantom loaded with Tc-99m are
obtained without and with material filter.
The performance of the tomographic system is explored via the uniformity,
sensitivity and linearity measurements, as well as the image lesi9n (hot) contrast,
signal-to-noise ratio, standard deviation and percent root mean square noise
(PRMSN). From the results, material filtered data shows an improvement in all
tested parameters at the cost of system sensitivity. |
format |
Thesis |
qualification_level |
Master's degree |
author |
Vee, Fong Shu |
author_facet |
Vee, Fong Shu |
author_sort |
Vee, Fong Shu |
title |
Uniformity, sensitivity, linearity, and hot region image analysis by employing unconventional filter technique (Cu 0.125mm and AI 0.2mm) with Tc-99m inSPECT imaging |
title_short |
Uniformity, sensitivity, linearity, and hot region image analysis by employing unconventional filter technique (Cu 0.125mm and AI 0.2mm) with Tc-99m inSPECT imaging |
title_full |
Uniformity, sensitivity, linearity, and hot region image analysis by employing unconventional filter technique (Cu 0.125mm and AI 0.2mm) with Tc-99m inSPECT imaging |
title_fullStr |
Uniformity, sensitivity, linearity, and hot region image analysis by employing unconventional filter technique (Cu 0.125mm and AI 0.2mm) with Tc-99m inSPECT imaging |
title_full_unstemmed |
Uniformity, sensitivity, linearity, and hot region image analysis by employing unconventional filter technique (Cu 0.125mm and AI 0.2mm) with Tc-99m inSPECT imaging |
title_sort |
uniformity, sensitivity, linearity, and hot region image analysis by employing unconventional filter technique (cu 0.125mm and ai 0.2mm) with tc-99m inspect imaging |
granting_institution |
Universiti Sains Malaysia |
granting_department |
Pusat Pengajian Sains Kesihatan |
publishDate |
2004 |
url |
http://eprints.usm.my/44154/1/LI...Fong%20Shu%20Tee...2004...-24%20pages.pdf |
_version_ |
1747821335570022400 |