Study on the effect of the sonication power and intersonication cooling time on the near field temperature in magnetic resonance high intensity focused ultrasound
Background: MR-HIFU is a technological breakthrough for non-surgical approach for treating tumour. MR-HIFU is safe with many advantages however complications may occur, at target and the near-field region. Specific ultrasound setup at MR-HIFU console must be made prior to ablation process, which...
Saved in:
Main Author: | |
---|---|
Format: | Thesis |
Language: | English |
Published: |
2018
|
Subjects: | |
Online Access: | http://eprints.usm.my/46513/1/Dr.%20Mohd%20Radhwan%20Abidin-24%20pages.pdf |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
id |
my-usm-ep.46513 |
---|---|
record_format |
uketd_dc |
spelling |
my-usm-ep.465132020-03-17T08:48:01Z Study on the effect of the sonication power and intersonication cooling time on the near field temperature in magnetic resonance high intensity focused ultrasound 2018 Abidin, Mohd Radhwan R Medicine (General) Background: MR-HIFU is a technological breakthrough for non-surgical approach for treating tumour. MR-HIFU is safe with many advantages however complications may occur, at target and the near-field region. Specific ultrasound setup at MR-HIFU console must be made prior to ablation process, which can affect the near-field temperature. The purpose of this study was to investigate how intersonication cooling time and sonication power influence the near-field temperature. Methodology: Philips QA phantom was used for ablation. A total of 100 ablations were made with different intersonication cooling time, which are 30 s, 60 s, 90 s and 120 s and different sonication power, which are 40 W, 80 W, 120 W, 160 W and 200 W. The near-field region was taken at 2cm from the base of phantom. The data were evaluated using statistical analysis. Level of significance was determined (p < 0.05). Results: There was no significant correlation between intersonication cooling time and the near-field temperature, with p = 0.089. However, there is a significant correlation between sonication power and the near-field temperature with p = 0.001. Regression analysis indicates that the sonication power significantly predicted the temperature changes [R2 = .416, F(1,18) = 14.533, p = .001] Conclusion: There is no relationship between intersonication cooling time with near-field temperature. However, there is relationship between sonication power and near-field temperature. 2018 Thesis http://eprints.usm.my/46513/ http://eprints.usm.my/46513/1/Dr.%20Mohd%20Radhwan%20Abidin-24%20pages.pdf application/pdf en public masters Universiti Sains Malaysia Pusat Pengajian Sains Perubatan |
institution |
Universiti Sains Malaysia |
collection |
USM Institutional Repository |
language |
English |
topic |
R Medicine (General) |
spellingShingle |
R Medicine (General) Abidin, Mohd Radhwan Study on the effect of the sonication power and intersonication cooling time on the near field temperature in magnetic resonance high intensity focused ultrasound |
description |
Background: MR-HIFU is a technological breakthrough for non-surgical approach for
treating tumour. MR-HIFU is safe with many advantages however complications may occur,
at target and the near-field region. Specific ultrasound setup at MR-HIFU console must be
made prior to ablation process, which can affect the near-field temperature. The purpose of
this study was to investigate how intersonication cooling time and sonication power influence
the near-field temperature.
Methodology: Philips QA phantom was used for ablation. A total of 100 ablations were
made with different intersonication cooling time, which are 30 s, 60 s, 90 s and 120 s and
different sonication power, which are 40 W, 80 W, 120 W, 160 W and 200 W. The near-field
region was taken at 2cm from the base of phantom. The data were evaluated using statistical
analysis. Level of significance was determined (p < 0.05).
Results: There was no significant correlation between intersonication cooling time and the
near-field temperature, with p = 0.089. However, there is a significant correlation between
sonication power and the near-field temperature with p = 0.001. Regression analysis indicates
that the sonication power significantly predicted the temperature changes [R2 = .416, F(1,18)
= 14.533, p = .001]
Conclusion: There is no relationship between intersonication cooling time with near-field
temperature. However, there is relationship between sonication power and near-field
temperature. |
format |
Thesis |
qualification_level |
Master's degree |
author |
Abidin, Mohd Radhwan |
author_facet |
Abidin, Mohd Radhwan |
author_sort |
Abidin, Mohd Radhwan |
title |
Study on the effect of the sonication power and intersonication cooling time on the near field temperature in magnetic resonance high intensity focused ultrasound |
title_short |
Study on the effect of the sonication power and intersonication cooling time on the near field temperature in magnetic resonance high intensity focused ultrasound |
title_full |
Study on the effect of the sonication power and intersonication cooling time on the near field temperature in magnetic resonance high intensity focused ultrasound |
title_fullStr |
Study on the effect of the sonication power and intersonication cooling time on the near field temperature in magnetic resonance high intensity focused ultrasound |
title_full_unstemmed |
Study on the effect of the sonication power and intersonication cooling time on the near field temperature in magnetic resonance high intensity focused ultrasound |
title_sort |
study on the effect of the sonication power and intersonication cooling time on the near field temperature in magnetic resonance high intensity focused ultrasound |
granting_institution |
Universiti Sains Malaysia |
granting_department |
Pusat Pengajian Sains Perubatan |
publishDate |
2018 |
url |
http://eprints.usm.my/46513/1/Dr.%20Mohd%20Radhwan%20Abidin-24%20pages.pdf |
_version_ |
1747821686178185216 |