The determination of absorbed dose in femur bone marrow during x-ray radiography through fabrication of physical and digital femur phantoms
The Monte Carlo simulation codes, BEAMnrc, were used to simulate a diagnostic Xray generator. First and second HVL with particle spectrums of three X-ray beams, viz. 84, 103 and 127 kVp were obtained using BEAMnrc simulation, IPEM report SpekCal 78, and measurements. The results of BEAMnrc simula...
محفوظ في:
المؤلف الرئيسي: | Dehyagani, Zaker Salehi |
---|---|
التنسيق: | أطروحة |
اللغة: | English |
منشور في: |
2013
|
الموضوعات: | |
الوصول للمادة أونلاين: | http://eprints.usm.my/60903/1/ZAKER%20SALEHI%20DEHYAGANI%20-%20e.pdf |
الوسوم: |
إضافة وسم
لا توجد وسوم, كن أول من يضع وسما على هذه التسجيلة!
|
مواد مشابهة
-
Bone mineral density in patient admitted to Hospital Raja Perempuan Zainab II with proximal femur fracture
بواسطة: Abd Wahab, Noor Hidayah Abdullah@
منشور في: (2016) -
Outcome of neck of femur fracture in elderly treated non operatively in Hospital Selayang
بواسطة: Appadurai, Shri Kapilan
منشور في: (2016) -
Measurement and calculated dose in breast phantom
بواسطة: Rahman, Siti Hajar Abdul
منشور في: (2004) -
Study of anthropometric measurement of distal femur among Malay, Chinese and Indian
بواسطة: Wei, Chew Yu
منشور في: (2021) -
Association of posterior cruciate ligament injury in different types of femur racture in Hospital Alor Setar Kedah, a cross sectional study
بواسطة: Singh, Inderjeet Singh A/L Ahtar
منشور في: (2016)