Measurement and calculated dose in breast phantom

The aim of this study was to compare the dose calculated by NUCLETRON PLATO treatment planning computer with thermoluminescence dosimeter (TLD) measurement in breast cancer radiotherapy. The absorbed dose distribution for one-field technique and two-field techniques were discussed. The study was...

全面介紹

Saved in:
書目詳細資料
主要作者: Rahman, Siti Hajar Abdul
格式: Thesis
語言:English
出版: 2004
主題:
在線閱讀:http://eprints.usm.my/46483/1/LI...Siti%20Hajar%20Binti%20Abdul%20Rahman...2004...-24%20pages.pdf
標簽: 添加標簽
沒有標簽, 成為第一個標記此記錄!
實物特徵
總結:The aim of this study was to compare the dose calculated by NUCLETRON PLATO treatment planning computer with thermoluminescence dosimeter (TLD) measurement in breast cancer radiotherapy. The absorbed dose distribution for one-field technique and two-field techniques were discussed. The study was done by comparing interest point dose values calculated by treatment planning computer with dose values measured by LiF TLDs. Utilizing an anthropomorphic phantom, TLDs were placed inside a breast phantom and were irradiated with 6MV SIEMENS MEVATRON linear accelerator. One-field technique and two-field techniques were used for the irradiation. Readings from irradiated TLDs were analyzed. Absorbed dose distributions for one-field and two-field techniques were obtained from the computer. Based on the results, dose values for the four points of interest in which calculated by the treatment planning computer and measured by TLDs were within 13%. Inexact location of points of measurement during phantom's set up caused large deviation between the calculation and TLD measurement. TLD measurement and the calculation do not comply to ICRU recommendation.