Quantitative Accuracy Of Different Single Photon Emission Computed Tomography/Computed Tomography (Spect/Ct) Systems With Phantoms Using Tc-99m And I-131
This project aimed to evaluate the performance of three SPECT/CT systems installed at three centres: the Advanced Medical and Dental Institute (AMDIUSM), Hospital Universiti Sains Malaysia (HUSM), and Penang General Hospital (PGH), and to assess levels of error in the quantitation of activity at th...
Saved in:
主要作者: | |
---|---|
格式: | Thesis |
语言: | English |
出版: |
2023
|
主题: | |
在线阅读: | http://eprints.usm.my/60539/1/Pages%20from%20MOAYED%20YOUSEF%20HASAN%20AL-QINNAH%20-%20TESIS.pdf |
标签: |
添加标签
没有标签, 成为第一个标记此记录!
|
id |
my-usm-ep.60539 |
---|---|
record_format |
uketd_dc |
spelling |
my-usm-ep.605392024-05-05T04:03:40Z Quantitative Accuracy Of Different Single Photon Emission Computed Tomography/Computed Tomography (Spect/Ct) Systems With Phantoms Using Tc-99m And I-131 2023-06 Al-Qinnah, Moayed Yousef Hasan R5-920 Medicine (General) This project aimed to evaluate the performance of three SPECT/CT systems installed at three centres: the Advanced Medical and Dental Institute (AMDIUSM), Hospital Universiti Sains Malaysia (HUSM), and Penang General Hospital (PGH), and to assess levels of error in the quantitation of activity at the centres involved. Furthermore, the aim is to evaluate the stability of the calibration factor of the SPECT/CT system at AMDI over a six-month period. 2023-06 Thesis http://eprints.usm.my/60539/ http://eprints.usm.my/60539/1/Pages%20from%20MOAYED%20YOUSEF%20HASAN%20AL-QINNAH%20-%20TESIS.pdf application/pdf en public phd doctoral Universiti Sains Malaysia Institut Perubatan & Pergigian Termaju (Advanced Medical & Dental Institute (AMDI) |
institution |
Universiti Sains Malaysia |
collection |
USM Institutional Repository |
language |
English |
topic |
R5-920 Medicine (General) |
spellingShingle |
R5-920 Medicine (General) Al-Qinnah, Moayed Yousef Hasan Quantitative Accuracy Of Different Single Photon Emission Computed Tomography/Computed Tomography (Spect/Ct) Systems With Phantoms Using Tc-99m And I-131 |
description |
This project aimed to evaluate the performance of three SPECT/CT systems installed at three centres: the Advanced Medical and Dental Institute (AMDIUSM), Hospital Universiti Sains Malaysia (HUSM), and Penang General Hospital
(PGH), and to assess levels of error in the quantitation of activity at the centres involved. Furthermore, the aim is to evaluate the stability of the calibration factor of the SPECT/CT system at AMDI over a six-month period. |
format |
Thesis |
qualification_name |
Doctor of Philosophy (PhD.) |
qualification_level |
Doctorate |
author |
Al-Qinnah, Moayed Yousef Hasan |
author_facet |
Al-Qinnah, Moayed Yousef Hasan |
author_sort |
Al-Qinnah, Moayed Yousef Hasan |
title |
Quantitative Accuracy Of Different Single Photon Emission Computed Tomography/Computed Tomography (Spect/Ct) Systems With Phantoms Using Tc-99m And I-131 |
title_short |
Quantitative Accuracy Of Different Single Photon Emission Computed Tomography/Computed Tomography (Spect/Ct) Systems With Phantoms Using Tc-99m And I-131 |
title_full |
Quantitative Accuracy Of Different Single Photon Emission Computed Tomography/Computed Tomography (Spect/Ct) Systems With Phantoms Using Tc-99m And I-131 |
title_fullStr |
Quantitative Accuracy Of Different Single Photon Emission Computed Tomography/Computed Tomography (Spect/Ct) Systems With Phantoms Using Tc-99m And I-131 |
title_full_unstemmed |
Quantitative Accuracy Of Different Single Photon Emission Computed Tomography/Computed Tomography (Spect/Ct) Systems With Phantoms Using Tc-99m And I-131 |
title_sort |
quantitative accuracy of different single photon emission computed tomography/computed tomography (spect/ct) systems with phantoms using tc-99m and i-131 |
granting_institution |
Universiti Sains Malaysia |
granting_department |
Institut Perubatan & Pergigian Termaju (Advanced Medical & Dental Institute (AMDI) |
publishDate |
2023 |
url |
http://eprints.usm.my/60539/1/Pages%20from%20MOAYED%20YOUSEF%20HASAN%20AL-QINNAH%20-%20TESIS.pdf |
_version_ |
1804888959594004480 |